T1 - Reliability of ITV approach to varying treatment fraction time: A retrospective analysis based on 2D cine MR images
AU - Cusumano, Davide
AU - Dhont, Jennifer
AU - Dhont, Jennifer
AU - Boldrini, Luca
AU - Chiloiro, Giuditta
AU - Romano, Angela
AU - Votta, Claudio
AU - Longo, Silvia
AU - Placidi, Lorenzo
AU - Azario, Luigi
AU - De Spirito, Marco
AU - Verellen, Dirk
AU - Valentini, Vincenzo
PY - 2020
Y1 - 2020
N2 - Background: Internal Target Volume (ITV) is one of the most common strategies to passively manage tumour motion in Radiotherapy (RT). The reliability of this approach is based on the assumption that the tumour motion estimated during pre-treatment 4D Computed Tomography (CT) acquisition is representative of the motion during the whole RT treatment. With the introduction of Magnetic Resonance-guided RT (MRgRT), it has become possible to monitor tumour motion during the treatment and verify this assumption. Aim of this study was to investigate the reliability of the ITV approach with respect to the treatment fraction time (TFT) in abdominal and thoracic lesions. Methods: A total of 12 thoracic and 15 abdominal lesions was analysed. Before treatment, a 10-phase 4DCT was acquired and ITV margins were estimated considering the envelope of the lesion contoured on the different 4DCT phases. All patients underwent MRgRT treatment in free-breathing, monitoring the tumour position on a sagittal plane with 4 frames per second (sec). ITV margins were projected on the tumour trajectory and the percentage of treatment time in which the tumour was inside the ITV (%TT) was measured to varying of TFT. The ITV approach was considered moderately reliable when %TT ≥ 90% and strongly reliable when %TT ≥ 95%. Additional ITV margins required to achieve %TT ≥ 95% were also calculated. Results: In the analysed cohort of patients, ITV strategy can be considered strongly reliable only for lung lesions with TFT ≤ 7 min (min). The ITV strategy can be considered only moderately reliable for abdominal lesions, and additional margins are required to obtain %TT ≥ 95%. Considering a TFT ≤ 4 min, additional margins of 2 mm in cranio-caudal (CC) and 1 mm in antero-posterior (AP) are suggested for pancreatic lesions, 3 mm in CC and 2 mm in AP for renal and liver ones. Conclusions: On the basis of the analysed cases, the ITV approach appears to be reliable in the thorax, while it results more challenging in the abdomen, due to the higher uncertainty in ITV definition and to the observed larger intra and inter-fraction motion variability. The addition of extra margins based on the TFT may represent a valid tool to compensate such limitations.
AB - Background: Internal Target Volume (ITV) is one of the most common strategies to passively manage tumour motion in Radiotherapy (RT). The reliability of this approach is based on the assumption that the tumour motion estimated during pre-treatment 4D Computed Tomography (CT) acquisition is representative of the motion during the whole RT treatment. With the introduction of Magnetic Resonance-guided RT (MRgRT), it has become possible to monitor tumour motion during the treatment and verify this assumption. Aim of this study was to investigate the reliability of the ITV approach with respect to the treatment fraction time (TFT) in abdominal and thoracic lesions. Methods: A total of 12 thoracic and 15 abdominal lesions was analysed. Before treatment, a 10-phase 4DCT was acquired and ITV margins were estimated considering the envelope of the lesion contoured on the different 4DCT phases. All patients underwent MRgRT treatment in free-breathing, monitoring the tumour position on a sagittal plane with 4 frames per second (sec). ITV margins were projected on the tumour trajectory and the percentage of treatment time in which the tumour was inside the ITV (%TT) was measured to varying of TFT. The ITV approach was considered moderately reliable when %TT ≥ 90% and strongly reliable when %TT ≥ 95%. Additional ITV margins required to achieve %TT ≥ 95% were also calculated. Results: In the analysed cohort of patients, ITV strategy can be considered strongly reliable only for lung lesions with TFT ≤ 7 min (min). The ITV strategy can be considered only moderately reliable for abdominal lesions, and additional margins are required to obtain %TT ≥ 95%. Considering a TFT ≤ 4 min, additional margins of 2 mm in cranio-caudal (CC) and 1 mm in antero-posterior (AP) are suggested for pancreatic lesions, 3 mm in CC and 2 mm in AP for renal and liver ones. Conclusions: On the basis of the analysed cases, the ITV approach appears to be reliable in the thorax, while it results more challenging in the abdomen, due to the higher uncertainty in ITV definition and to the observed larger intra and inter-fraction motion variability. The addition of extra margins based on the TFT may represent a valid tool to compensate such limitations.
KW - Abdominal Neoplasms
KW - Artifacts
KW - Four-Dimensional Computed Tomography
KW - Humans
KW - Internal target volume
KW - Intra-fraction motion management
KW - MR-guided radiotherapy
KW - Magnetic Resonance Imaging
KW - Motion
KW - Radiosurgery
KW - Radiotherapy Planning, Computer-Assisted
KW - Radiotherapy, Image-Guided
KW - Retrospective Studies
KW - Thoracic Neoplasms
KW - Abdominal Neoplasms
KW - Artifacts
KW - Four-Dimensional Computed Tomography
KW - Humans
KW - Internal target volume
KW - Intra-fraction motion management
KW - MR-guided radiotherapy
KW - Magnetic Resonance Imaging
KW - Motion
KW - Radiosurgery
KW - Radiotherapy Planning, Computer-Assisted
KW - Radiotherapy, Image-Guided
KW - Retrospective Studies
KW - Thoracic Neoplasms
UR - http://hdl.handle.net/10807/207160