Rapid determination of short chain carnitines in human plasma by electrospray ionisation-ion trap mass spectrometry using capillary electrophoresis instrument as sampler.

Claudia Desiderio, A Mancinelli, A De Rossi, Diana Valeria Rossetti, Rosanna Inzitari, I Messana, Bruno Giardina, Massimo Castagnola

Risultato della ricerca: Contributo in rivistaArticolo in rivista

11 Citazioni (Scopus)

Abstract

A capillary electrophoresis apparatus was used as sampler for flow injection analysis (FIA) in tandem mass spectrometry Of L-Carnitine and its acetyl- and propionyl-metabolites in human plasma. The capillary electrophoresis instrument was coupled to the ion trap mass spectrometer by an electrospray ionization coaxial sheath liquid interface. The electrophoresis capillary introduced the sample directly into the source by applying a prolonged sample injection. The use of the capillary electrophoresis apparatus miniaturised the FIA procedure, substantially reducing the quantities of solvents and samples used, and allowed rapid automated sequential analyses. The method was optimised and validated. using a dialyzed human plasma matrix. The plasma samples were analysed after a simple, rapid deproteinisation procedure with acetonitrile and diluted 70 times before direct injection into the mass spectrometer for product ion scan MS/MS analysis in positive ionisation. The total analysis time was 5 min, including capillary preconditioning and acquisition time (3 min). The method was sensitive, allowing the determination Of L-, L-acetyl- and L-propionyl-carnitines at 140, 14 and 3.6 nM concentrations (injected values) corresponding to lower limit of quantitation values in plasma of 10, 1 and 0.25 mu M, respectively. The method was processed for full validation and applied to the analysis Of L-carnitine and its short chain derivatives in human plasma samples. (C) 2006 Elsevier B.V. All rights reserved.
Lingua originaleEnglish
pagine (da-a)320-326
Numero di pagine7
RivistaJournal of Chromatography A
Volume1150
Stato di pubblicazionePubblicato - 2007

Keywords

  • capillary
  • electrophoresis

Fingerprint

Entra nei temi di ricerca di 'Rapid determination of short chain carnitines in human plasma by electrospray ionisation-ion trap mass spectrometry using capillary electrophoresis instrument as sampler.'. Insieme formano una fingerprint unica.

Cita questo