TY - JOUR
T1 - Possible Radio Interference between Video Capsule Endoscopy and Second-Generation OmniPod Patch Pump
AU - Pitocco, Dario
AU - Rizzi, Alessandro
AU - Tortora, Annalisa
AU - Manto, Andrea
AU - Zaccardi, Francesco
AU - Ghirlanda, Giovanni
AU - Costamagna, Guido
AU - Riccioni, Maria Elena
PY - 2016
Y1 - 2016
N2 - Video capsule endoscopy (VCE) is a noninvasive diagnostic tool used to observe the small intestinal mucosa. We report a case of a 57-year-old woman with T2DM, treated with continuous subcutaneous insulin infusion using second-generation OmniPod patch pump, undergoing VCE (Given M2A; VCE Ltd, Yoqneam, Israel) for melena and anemia. During VCE, an abnormal interruption of communication between video capsule and its receiver occurred. Two hours after capsule ingestion, the patient activated the insulin pump infusion through the Personal Diabetes Manager (PDM) because she drank a sugary beverage for the first time after ingestion. Due to this, we decided to repeat VCE after the removal of the insulin pump and PDM: at this time, the capsule recorded for more than 10 h without any interruption. The video capsule and second-generation OmniPod patch pump use the same radio frequency and this may cause interference between these two devices. In patients using second-generation OmniPod patch pump undergoing VCE, we suggest to switch to intravenous insulin infusion or multiple daily injection or to use a different model of VCE, as MiRoCam (Intromedic, Seoul, Korea).
AB - Video capsule endoscopy (VCE) is a noninvasive diagnostic tool used to observe the small intestinal mucosa. We report a case of a 57-year-old woman with T2DM, treated with continuous subcutaneous insulin infusion using second-generation OmniPod patch pump, undergoing VCE (Given M2A; VCE Ltd, Yoqneam, Israel) for melena and anemia. During VCE, an abnormal interruption of communication between video capsule and its receiver occurred. Two hours after capsule ingestion, the patient activated the insulin pump infusion through the Personal Diabetes Manager (PDM) because she drank a sugary beverage for the first time after ingestion. Due to this, we decided to repeat VCE after the removal of the insulin pump and PDM: at this time, the capsule recorded for more than 10 h without any interruption. The video capsule and second-generation OmniPod patch pump use the same radio frequency and this may cause interference between these two devices. In patients using second-generation OmniPod patch pump undergoing VCE, we suggest to switch to intravenous insulin infusion or multiple daily injection or to use a different model of VCE, as MiRoCam (Intromedic, Seoul, Korea).
KW - Endocrinology
KW - Endocrinology, Diabetes and Metabolism
KW - Medical Laboratory Technology
KW - Endocrinology
KW - Endocrinology, Diabetes and Metabolism
KW - Medical Laboratory Technology
UR - http://hdl.handle.net/10807/93145
UR - http://www.liebertonline.com/dia
U2 - 10.1089/dia.2016.0007
DO - 10.1089/dia.2016.0007
M3 - Article
SN - 1520-9156
VL - 18
SP - 444
EP - 445
JO - DIABETES TECHNOLOGY & THERAPEUTICS
JF - DIABETES TECHNOLOGY & THERAPEUTICS
ER -