Multicenter evaluation of an enzymatic method for glycated albumin

  • Renata Paleari
  • , Graziella Bonetti
  • , Cinzia Anna Maria Calla'
  • , Mariarosa Carta
  • , Ferruccio Ceriotti
  • , Nicola Di Gaetano
  • , Marilisa Ferri
  • , Elena Guerra
  • , Gabriella Lavalle
  • , Claudia Lo Cascio
  • , Francesca Gabriela Martino
  • , Martina Montagnana
  • , Marco Moretti
  • , Gabriele Santini
  • , Donata Scribano
  • , Roberto Testa
  • , Anna Vero
  • , Andrea Mosca
  • University of Milan
  • Spedali Civili Di Brescia
  • Azienda Sanitaria Ulss 6 Vicenza
  • IRCCS Fondazione Ca'Granda – Ospedale Maggiore Policlinico - Milano
  • Werfen
  • Laboratorio Analisi di Urbino
  • San Raffaele Scientific Institute
  • Ospedale di Bracciano
  • Ospedale Policlinico
  • U.O.C. Patologia Clinica (HB 5 Regionale) P.O. San Filippo Neri ASL ROMA A
  • University of Verona
  • Patologia Clinica Azienda Ospedaliera Ospedali Riuniti Marche Nord Pesaro
  • Istituto Nazionale di Riposo e Cura per Anziani - Ancona
  • Azienda Ospedaliera Pugliese-Ciaccio

Risultato della ricerca: Contributo in rivista › Articolo › peer review

Abstract

BACKGROUND: The use of glycated albumin (GA) has been proposed as an additional glycemic control marker particularly useful in intermediate-term monitoring and in situation when HbA1c test is not reliable. METHODS: We have performed the first multicenter evaluation of the analytical performance of the enzymatic method quantILab Glycated Albumin assay implemented on the most widely used clinical chemistry analyzers (i.e. Abbott Architect C8000, Beckman Coulter AU 480 and 680, Roche Cobas C6000, Siemens ADVIA 2400 and 2400 XPT). RESULTS: The repeatability of the GA measurement (expressed as CV, %) implemented in the participating centers ranged between 0.9% and 1.2%. The within-laboratory CVs ranged between 1.2% and 1.6%. A good alignment between laboratories was found, with correlation coefficients from 0.996 to 0.998. Linearity was confirmed in the range from 7.6 to 84.7%. CONCLUSION: The new enzymatic method for glycated albumin evaluated by our investigation is suitable for clinical use.
Lingua originaleInglese
pagine (da-a)81-86
Numero di pagine6
RivistaClinica Chimica Acta
Volume2017
DOI
Stato di pubblicazionePubblicato - 2017

OSS delle Nazioni Unite

Questo processo contribuisce al raggiungimento dei seguenti obiettivi di sviluppo sostenibile

  1. SDG 3 - Salute e benessere
    SDG 3 Salute e benessere

Keywords

  • Glicated Albumin
  • Multicenter evaluation
  • diabetes

Fingerprint

Entra nei temi di ricerca di 'Multicenter evaluation of an enzymatic method for glycated albumin'. Insieme formano una fingerprint unica.

Cita questo