Background: Extended half-life recombinant FVIII products (EHL-rFVIIIs) have been engineered to improve the pharmacokinetic profile of FVIII, enabling better hemostatic protection with a reduced number of injections in persons with hemophilia. Previous studies showed several discrepancies in FVIII activity (FVIII:C) measurements for EHL-rFVIIIs comparing one-stage clotting assay (OSA) and chromogenic assay (CSA), although a systematic investigation of this phenomenon is still lacking. Objective: Evaluation of the accuracy and precision of measurement of all available EHL-rFVIIIs with 5 different assays both in vitro and ex vivo. Methods: Damoctocog alfa pegol, rurioctocog alfa pegol, turoctocog alfa pegol, and efmoroctocog alfa were tested with 3 OSA types: (1) aPTT-based commercial reagents with colloidal silica (Synthasil, Werfen-IL); (2) ellagic acid, Synthafax (Werfen-IL); and (3) OSA calibrated with each EHL-rFVIII product and colloidal silica. Measurements were also carried out with 2 different commercially available CSA reagents (Coamatic Factor VIII, Chromogenix-Werfen) and Trinichrom FVIII (Tcoag-Stago). A Bland-Altman analysis was performed to compare all assays. Results: The simple OSA showed significant discrepancies between the expected and measured EHL-rFVIII concentrations as CSA methods, whereas the calibrated OSA assay was accurate and precise in determining the activity of all EHL-rFVIIIs in the in vitro setting. Comparable results were found using ex vivo plasma samples. Conclusion: In this study, only OSA with a calibration curve constructed with each EHL-rFVIII product showed acceptable accuracy and precision in EHL-rFVIIIs measurements.
| Lingua originale | Inglese |
|---|---|
| pagine (da-a) | 100070-100080 |
| Numero di pagine | 11 |
| Rivista | Research and Practice in Thrombosis and Haemostasis |
| Volume | 7 |
| DOI | |
| Stato di pubblicazione | Pubblicato - 2023 |
Keywords
- blood coagulation
- coagulation factor VIII
- congenital
- extended half-life rFVIII
- factor VIII deficiency
- tests
- therapeutic drug monitoring
Fingerprint
Entra nei temi di ricerca di 'Measurement of extended half-life recombinant FVIII molecules: In vitro and ex vivo evidence of relevant assay discrepancies'. Insieme formano una fingerprint unica.Cita questo
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver