TY - JOUR
T1 - Biomechanical evaluation of different balloon positions for proximal optimization technique in left main bifurcation stenting
AU - Rigatelli, Gianluca
AU - Zuin, Marco
AU - Chiastra, Claudio
AU - Burzotta, Francesco
PY - 2020
Y1 - 2020
N2 - Background: Proximal optimization technique (POT) is a key step during left main (LM) bifurcation stenting. However, after crossover stenting, the ideal position of POT balloon is unclear. We sought to evaluate the biomechanical impact of different POT balloon positions during LM cross-over stenting procedure. Methods: We reconstructed the patient-specific LM bifurcation anatomy, using coronary computed tomography angiography data of 5 consecutive patients (3 males, mean age 66.3 ± 21.6 years) with complex LM bifurcation disease, defined as Medina 1,1,1, evaluated between 1st January 2018 to 1st June 2018 at our center. Finite element analyses were carried out to virtually perform the stenting procedure. POT was virtually performed in a mid (marker just at the carina cut plane), proximal (distal marker 1 mm before the carina) and distal (distal marker 1 mm after the carina) position in each investigated case. Final left circumflex obstruction (SBO%), strut malapposition, elliptical ratio and stent malapposition were evaluated. Results: The use of both proximal and distal POT resulted in a smaller LM diameter compared to the mid POT. SBO was significantly higher in both proximal and distal configurations compared to mid POT: 38.3 ± 5.1 and 29.3 ± 3.1 versus 18.3 ± 3.6%, respectively. Similarly stent malapposition was higher in both proximal and distal configurations compared to mid POT: 1.3 ± 0.4 and 0.82 ± 1.8 versus 0.78 ± 1.2, respectively. Conclusions: Mid POT offers the best results in terms of LCx opening maintaining slightly smaller but still acceptable LM and LAD diameters compared to alternative POT configuration.
AB - Background: Proximal optimization technique (POT) is a key step during left main (LM) bifurcation stenting. However, after crossover stenting, the ideal position of POT balloon is unclear. We sought to evaluate the biomechanical impact of different POT balloon positions during LM cross-over stenting procedure. Methods: We reconstructed the patient-specific LM bifurcation anatomy, using coronary computed tomography angiography data of 5 consecutive patients (3 males, mean age 66.3 ± 21.6 years) with complex LM bifurcation disease, defined as Medina 1,1,1, evaluated between 1st January 2018 to 1st June 2018 at our center. Finite element analyses were carried out to virtually perform the stenting procedure. POT was virtually performed in a mid (marker just at the carina cut plane), proximal (distal marker 1 mm before the carina) and distal (distal marker 1 mm after the carina) position in each investigated case. Final left circumflex obstruction (SBO%), strut malapposition, elliptical ratio and stent malapposition were evaluated. Results: The use of both proximal and distal POT resulted in a smaller LM diameter compared to the mid POT. SBO was significantly higher in both proximal and distal configurations compared to mid POT: 38.3 ± 5.1 and 29.3 ± 3.1 versus 18.3 ± 3.6%, respectively. Similarly stent malapposition was higher in both proximal and distal configurations compared to mid POT: 1.3 ± 0.4 and 0.82 ± 1.8 versus 0.78 ± 1.2, respectively. Conclusions: Mid POT offers the best results in terms of LCx opening maintaining slightly smaller but still acceptable LM and LAD diameters compared to alternative POT configuration.
KW - Bifurcations
KW - Left main
KW - POT
KW - Bifurcations
KW - Left main
KW - POT
UR - http://hdl.handle.net/10807/158781
U2 - 10.1016/j.carrev.2020.05.028
DO - 10.1016/j.carrev.2020.05.028
M3 - Article
SN - 1553-8389
SP - 30300
EP - 30306
JO - Cardiovascular Revascularization Medicine
JF - Cardiovascular Revascularization Medicine
ER -