Abstract
OBJECTIVE: To compare helmet noninvasive ventilation (NIV), in terms of
patient-ventilator interaction and performance, using 2 different circuits for
connection: a double tube circuit (with one inspiratory and one expiratory line)
and a standard circuit (a Y-piece connected only to one side of the helmet,
closing the other side).
METHODS: A manikin, connected to a test lung set at 2 breathing frequencies (20
and 30 breaths/min), was ventilated in pressure support ventilation (PSV) mode
with 2 different settings, randomly applied, of the ratio of pressurization time
to expiratory trigger time (T(press)/T(exp-trigger)) 50%/25%, default setting,
and T(press)/T(exp-trigger) 80%/60%, fast setting, through a helmet. The helmet
was connected to the ventilator randomly with the double and the standard
circuit. We measured inspiratory trigger delay (T(insp-delay)), expiratory
trigger delay (T(exp-delay)), T(press)), time of synchrony (T(synch)), trigger
pressure drop, inspiratory pressure-time product (PTP), PTP at 300 ms and 500 ms,
and PTP at 500 ms expressed as percentage of an ideal PTP500 (PTP500 index).
RESULTS: At both breathing frequencies and ventilator settings, helmet NIV with
the double tube circuit showed better patient-ventilator interaction, with
shorter T(insp-delay), T(exp-delay), and T(press); longer T(synch); and higher
PTP300, PTP500, and PTP500 index (all P < .01).
CONCLUSIONS: The double tube circuit had significantly better patient-ventilator
interaction and a lower rate of wasted effort at 30 breaths/min.
| Lingua originale | Inglese |
|---|---|
| pagine (da-a) | 1474-1481 |
| Numero di pagine | 8 |
| Rivista | Respiratory Care |
| Volume | 2013 |
| DOI | |
| Stato di pubblicazione | Pubblicato - 2013 |
Keywords
- noninvasive ventilation
Fingerprint
Entra nei temi di ricerca di 'A bench study of 2 ventilator circuits during helmet noninvasive ventilation'. Insieme formano una fingerprint unica.Cita questo
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver