Abstract
Reduced blood flow of from 43 to 71% has been reported in sutured and stapled anastomoses. The sutureless sliding absorbable, intraluminal, nontoxic stent (SAINT)-fibrin glue anastomotic method, which clamps the stump margins between 2 dissolving surfaces, includes only two stages of temporary compression labour 6 min total using 4 IU/mL thrombin) during the glue application in order to promote vascularization. A SAINT placement device (SAINT-PD) was introduced to facilitate low rectal anastomoses. Morphohistologic results from limited trials using fibrin glue with an untied sutureless stapler technique and a prototype non-gear-driven SAINT-PD, neither having the two dissolvable clamping surfaces of the SAINT, showed a 29 and 25% incidence of intraluminal tissue ridges, respectively Since these tissue ridges could result in subclinical dilatation or frank stenosis, and the more extensive SAINT trials had an 8% incidence of tissue ridges, redesign of the SAINT-PD was undertaken. Consequently, to improve the anastomotic quality of the SAINT-PD, the sliding absorbable reinforced ring (sucrose base) acting as the second dissolvable surface for the SAINT-PD and a new axially controlled geared SAINT-PD design are described.
| Original language | English |
|---|---|
| Pages (from-to) | 495-504 |
| Number of pages | 10 |
| Journal | Journal of Investigative Surgery |
| Volume | 9 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - 1996 |
All Science Journal Classification (ASJC) codes
- Surgery
Keywords
- Absorbable stent
- Anastomosis
- Fibrin glue
- Gastrointestinal tract
- Ischemia
- Stent placement device
- Swine
- Surgical
- Animals
- Digestive System
- Digestive System Surgical Procedures
- Fibrin Tissue Adhesive
- Materials Testing
- Stents
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver